RCYAGO DO9100

Meter Oxygène levona miaraka amin'ny fluid famenoana elektroda

Brand: RCYAGO | Model: DO9100

1. Fampidirana

The RCYAGO Dissolved Oxygen Meter is a high-precision instrument designed for accurate measurement of dissolved oxygen (DO) and temperature in various water bodies. It utilizes a polarographic sensor with a thermistor for automatic temperature compensation, ensuring reliable readings across different conditions. This manual provides comprehensive instructions for setup, operation, and maintenance of your DO meter.

RCYAGO Dissolved Oxygen Meter with probe

Figure 1: RCYAGO Dissolved Oxygen Meter and Probe

2. Inona no ao anaty boaty

Azafady, jereo ny votoatin'ny fonosana mba hahazoana antoka fa misy ny entana rehetra:

  • 1 x RCYAGO Dissolved Oxygen Meter (Model: DO9100)
  • 1 x Electrode Probe
  • 6 x Probe Protective Covers (Membrane Sleeves)
  • 1 x Electrode Filling Solution
  • 1 x Anaerobic Water Powder (for calibration)
  • 1 x User Manual

3. Fametrahana

3.1 Fametrahana bateria

The meter requires 3 x 1.5V button batteries. These are typically pre-installed or included in the top cap of the device. Ensure they are correctly oriented before use.

3.2 Fifandraisana electrode

Connect the electrode probe to the main meter unit by screwing the metal knob securely. Ensure a firm connection to prevent inaccurate readings.

Diagram showing dimensions, LCD display, power button, lock/temperature switch key, mode/calibration key, and electrode interface.

Figure 2: Meter Components and Dimensions

3.3 Filling the Probe Protective Cover

The probe protective cover contains a permeable membrane. Before first use, or when replacing the membrane sleeve, fill it with the provided electrode filling solution. This solution is crucial for the sensor's proper function. Do not press on the membrane itself.

Close-up of the highly sensitive probe with multiple probe protective covers shown. The image highlights the detachable nature of the probe and the included covers.

Figure 3: Highly Sensitive Probe and Protective Covers

4. Fampandehanana

4.1 Famelomana/Famonoana ary Fampisehoana

Tsindrio ny bokotra herinaratra () to turn the meter on. The digital screen will illuminate with a backlight, displaying DO readings and temperature. The backlight improves readability in low-light conditions. The device features an auto-power-off function after 8 minutes of inactivity to conserve battery life.

Close-up of the meter's HD backlight display showing readings, and an open battery compartment.

Figure 4: HD Backlight Display and Battery Compartment

4.2 Fomba fandrefesana

The meter can display dissolved oxygen in mg/L (parts per million, ppm) or as a saturation percentage (%). Press the MODE/CAL button to toggle between these modes.

4.3 Onitra amin'ny maripana

The meter features automatic temperature compensation (ATC) within the range of 0-50°C (32-122°F). This ensures that dissolved oxygen readings are accurate despite variations in water temperature, as oxygen solubility is affected by temperature.

Dissolved oxygen meter probe submerged in an aquarium, illustrating automatic temperature compensation.

Figure 5: Automatic Temperature Compensation in action

4.4 Kalibration

Regular calibration is essential for accurate measurements. The meter supports two-point calibration: 100% air calibration and 0% anaerobic water calibration.

4.4.1 100% Air Calibration

  1. Ensure the probe is exposed to air.
  2. Tsindrio lava ny MODE/CAL bokotra mandritra ny 5 segondra.
  3. Release the button. The screen will flash 3 times at 100%. Calibration is complete.

4.4.2 0% Anaerobic Water Calibration

  1. Prepare anaerobic water: Pour 1 packet of anaerobic powder into the attached calibration bottle, add 30ml of purified water or tap water, and wait 5-10 minutes for the powder to fully dissolve.
  2. Submerge the electrode in the anaerobic water.
  3. Tsindrio ny MODE/CAL key to switch to mg/L mode.
  4. Wait for the value to drop to 0.
  5. Tsindrio lava ny MODE/CAL bokotra mandritra ny 5 segondra.
  6. Release the button. The screen will flash 3 times at 0%. Calibration is complete.
Diagram illustrating the steps for 0% anaerobic water calibration and 100% air calibration.

Figure 6: Calibration Procedures

4.5 Mandray refy

Once calibrated, submerge the electrode probe into the water sample you wish to measure. The meter will display the dissolved oxygen and temperature readings. Allow a few moments for the readings to stabilize.

Person holding the dissolved oxygen meter, with the probe submerged in a pond with fish, showing a real-world application.

Figure 7: Measuring Dissolved Oxygen in a Pond

4.6 Switching Temperature Units

To switch between Celsius (°C) and Fahrenheit (°F), long press the HAZO/TEMP bokotra.

4.7 Understanding Oxygen Solubility

Oxygen solubility in water is influenced by temperature, atmospheric pressure, and salinity. The provided table illustrates the saturation concentration of oxygen at different temperatures:

Table showing saturation concentration of oxygen in mg/L and percentage at various temperatures in Celsius and Fahrenheit.

Figure 8: Oxygen Saturation Concentration at Different Temperatures

5. Fikojakojana

5.1 Probe Protective Cover (Membrane Sleeve)

The film inside the protective cover cannot be taken apart. If test values become inaccurate even after calibration, the membrane sleeve needs to be replaced. The meter comes with 6 spare covers.

5.2 Fanadiovana

The meter unit and probe should be cleaned with care. The product care instructions specify 'Hand Wash Only'. Avoid harsh chemicals or abrasive materials.

5.3 Fitehirizana

When not in use, ensure the probe protective cover is properly filled with solution and securely attached to protect the sensor. Store the meter in a cool, dry place.

6. Famahana olana

  • Famakiana tsy marina: Ensure the electrode is properly connected and the membrane sleeve is filled with the electrode filling fluid. Perform calibration regularly (both air and anaerobic water calibration) as described in Section 4.4.
  • Olana amin'ny fampisehoana: Check battery levels. If the display is dim or erratic, replace the batteries.
  • No Reading: Verify the electrode is fully submerged in the sample and that the connection to the meter is secure.
  • Ny tsy fahombiazan'ny calibration: Ensure the anaerobic water powder is fully dissolved for 0% calibration. For air calibration, ensure the probe is exposed to stable ambient air. If issues persist, the membrane sleeve might need replacement.

7. Famaritana

endri-javatraantsipirihany
Ny refin'ny vokatra6.1 x 0.01 x 0.01 santimetatra
Laharana modelyD9100
lanja12.56 oz
MarikaRCYAGO
Endri-javatra manokanaHigh Accuracy
lokoMainty
Fitaovana ivelanyplastika
Karazana fampisehoanaDigital with Backlight
Dissolved Oxygen Range0.0-40.0 mg/L
Dissolved Oxygen Saturation0.0-300.0%
Temperature Range0.0-40.0°C (32.0-104.0°F)
Onitra ho azy ny maripana0-50°C
Loharanon-kery3 x 1.5V Button Batteries
Auto fanakatonana8 minitra

8. Lahatsary vokatra

8.1 Digital Dissolved Oxygen Meter - Basic Operation & Calibration

Ity lahatsary ity dia manome topimaso fohyview of the RCYAGO Dissolved Oxygen Meter, demonstrating how to power it on, install the electrode, and perform both air and oxygen-free water calibrations. It highlights the key steps for initial setup and ensuring accurate readings.

8.2 Portable Dissolved Oxygen Meter - Detailed Review & Fampiasana

This comprehensive video offers a detailed review of the portable dissolved oxygen meter. It covers unboxing, battery installation, electrode setup, and practical demonstrations of measuring dissolved oxygen in different water samples (tap water vs. boiled water). The video also discusses the included accessories like membrane caps and electrolyte solution, and explains the temperature conversion chart.

9. Warranty sy fanohanana

For warranty information, technical support, or to purchase replacement parts (such as probe protective covers or electrode filling solution), please contact RCYAGO customer service. Refer to the product packaging or the official RCYAGO webtranokala ho an'ny antsipiriany momba ny fifandraisana.

Documents mifandraika - D9100

Preview DO-9100 Pen Type Intelligent Dissolved Oxygen Analyzer Instruction Manual
This document provides instructions for the DO-9100 pen type intelligent dissolved oxygen analyzer, covering its features, technical parameters, operation, maintenance, and troubleshooting.
Preview RCYAGO SGW07 Dual Water Timer Manual
Torolàlana momba ny fampiasana feno ho an'ny RCYAGO SGW07 Dual Water Timer. Mianara momba ny vokatraview, fepetra arahana, fametrahana, fandaharana fanondrahana mandeha ho azy, fanondrahana tanana, fanemorana ny orana, famahana olana, ary ny satan'ny bateria ho an'ny rafi-panondrahana zaridainanao.
Preview Портативный оксиметр Yinmik DO9100, BLE9100: Руководство пользователя
Полное руководство пользователя для портативного оксиметра Yinmik моделей DO9100 sy BLE9100. Охватывает технические характеристики, калибровку, эксплуатацию, обслуживание и гарантию.