APEX Domus 3

Apex Medical Domus 3 Alternating Pressure Mattress System User Manual

Model: Domus 3

1. Fampidirana

This manual provides comprehensive instructions for the safe and effective use of the Apex Medical Domus 3 Alternating Pressure Mattress System. This system is designed to assist in the prevention and treatment of pressure injuries, ulcers, and bedsores for individuals requiring extended bed rest. Please read this manual thoroughly before setup and operation.

Apex Medical Domus 3 Alternating Pressure Mattress System with pump on a hospital bed.

Sary 1.1: The Apex Medical Domus 3 Alternating Pressure Mattress System. This image displays the complete system, including the mattress installed on a hospital bed frame and the external pump unit. The mattress features individual air cells, and the pump is shown connected to the mattress.

2. Information momba ny fiarovana

Adherence to the following safety guidelines is crucial for patient well-being and proper system function.

3. Ireo singa

The Apex Medical Domus 3 system typically includes the following components:

4. Torolàlana fananganana

  1. Omano ny rafitry ny fandriana: Place the mattress system directly onto a standard hospital or home bed frame. Ensure the bed frame is clean and free of sharp objects that could damage the mattress.
  2. Apetraho ny kidoro: Unroll the Domus 3 mattress on the bed frame with the air hose connection end positioned towards the foot of the bed where the pump will be placed. Ensure the CPR strap is accessible.
  3. Mampifandray ny paompy: Attach the air hoses from the mattress to the corresponding connectors on the variable pressure pump unit. Ensure connections are secure.
  4. Mount the Pump: Hang the pump unit securely at the foot of the bed using its integrated hangers.
  5. Hery mampifandray: Plug the pump's power cord into a grounded electrical outlet.
  6. Fisondrotry ny vidim-piainana voalohany: Turn on the pump. The mattress will begin to inflate. Allow approximately 20-30 minutes for the mattress to fully inflate and stabilize before placing a patient on it.
Apex Medical Domus 3 mattress cover with zipper and water-resistant material detail.

Sary 4.1: Mattress Cover Features. An image showcasing two features of the mattress cover: a close-up of the zipper-all-around design for easy removal and replacement, and a magnified view of the Nylon/PU cover material demonstrating its water-resistant properties with water droplets on the surface.

5. Torolàlana miasa

5.1 Power On/off

5.2 Therapy Modes

The Domus 3 offers two therapy modes:

Refer to the pump unit's controls to select the desired therapy mode.

Diagram illustrating the alternating pressure redistribution cycle of the mattress.

Sary 5.1: Alternating Pressure Cycle. This diagram shows a human figure lying on the mattress, with an illustration of the alternating pressure cycle. Blue blocks represent inflated air cells, and arrows indicate the cyclical inflation and deflation pattern designed to redistribute pressure and prevent injury.

5.3 Pressure Adjustment (Firmness)

The pump unit features a variable pressure setting with a weight indicator.

5.4 Low Air Loss Function

Air cells 3 through 11 are designed with a low air loss feature. This function continuously ventilates air to wick moisture and heat away from the mattress surface, contributing to patient comfort and skin integrity. This operates automatically when the system is powered on.

Close-up of Apex Medical Domus 3 mattress showing low air loss ventilation.

Sary 5.2: Low Air Loss Feature. A detailed view of the mattress surface, highlighting the low air loss feature. Blue arrows indicate airflow through the air cells, designed to reduce moisture and heat for patient comfort and prevent skin maceration.

6. Fikojakojana sy fikarakarana

6.1 Cleaning the Mattress and Cover

6.2 Air Cell Replacement

Each of the 20 individual 8-inch air cells can be removed and replaced if damaged.

  1. Turn off the pump and disconnect it from the mattress.
  2. Unzip the mattress cover.
  3. Carefully detach the damaged air cell from its connections within the mattress structure.
  4. Install a new, compatible air cell, ensuring it is securely connected.
  5. Zip the mattress cover back into place and re-connect the pump for inflation.

7. Famahana olana

OLANAAntony mety hitrangavahaolana
Mattress not inflating or losing pressure
  • Tapaka herinaratra
  • Air hoses not securely connected
  • Leak in mattress or air cells
  • Tsy miasa ny paompy
  • Check power connection to outlet and pump.
  • Ensure air hoses are firmly attached to both mattress and pump.
  • Inspect mattress and individual air cells for punctures or damage. Replace damaged cells if necessary.
  • Contact customer support if pump appears to be malfunctioning.
Low pressure alarm sounds
  • Mattress not fully inflated
  • Setroka rivotra
  • Patient weight exceeds capacity
  • Allow more time for inflation.
  • Check for leaks in mattress or hose connections.
  • Verify patient weight is within the 440 lbs capacity.
Pump is unusually loud
  • Obstruction around pump fan
  • Tsy miasa ny paompy
  • Ensure adequate ventilation around the pump unit.
  • Contact customer support if noise persists. (Note: Pump noise level is typically ≤ 30dB(A)).

8. Famaritana

endri-javatraantsipirihany
MODELYDomus 3
Mattress Dimensions (L x W x H)78.7 x 35.4 x 8 santimetatra
Lanja rafitraManodidina ny 22.3 kilao
Weight Capacity440 kg
Karazana lambaNylon, Polyurethane (Cover)
Endri-javatra manokanaLow Air Loss, Alternating Pressure, Water Resistant Cover, CPR Strap, Low Pressure Alarm
Pump Noise Level≤ 30dB(A)
mpanamboatraFitsaboana Apex
Dimensions of Apex Medical Domus 3 mattress and pump system.

Sary 8.1: System Dimensions. A visual representation of the dimensions for both the Apex Medical Domus 3 mattress and its accompanying pump unit. Key measurements for length, width, and height are indicated for both components.

Apex Medical Domus mattress showing two height options, 5-inch and 8-inch.

Sary 8.2: Mattress Height Options. This image illustrates the two available height options for the Domus mattress: a 5-inch version and an 8-inch version. The current product is the 8-inch model, providing enhanced support.

9. Warranty sy fanohanana

9.1 Fampahalalana momba ny fiantohana

The Apex Medical Domus 3 Alternating Pressure Mattress System comes with a antoka 1 taona on both the mattress and the pump unit. This warranty covers manufacturing defects and malfunctions under normal use. Please retain your proof of purchase for warranty claims.

9.2 Fanohanana ny mpanjifa

For technical assistance, troubleshooting beyond this manual, or warranty inquiries, please contact Apex Medical customer support. Refer to the product packaging or the official Apex Medical webtranokala ho an'ny fampahalalana mifandraika amin'izao fotoana izao.

Documents mifandraika - Domus 3

Preview APEX XT Auto Provider Mode Quick Guide
A quick guide for medical professionals on how to use the APEX XT Auto device in Provider Mode, including setup navigation and resetting compliance meters.
Preview Torolàlana ho an'ny mpampiasa APEX VAC Pump: Torolàlana sy Torolàlana momba ny fiarovana
Torolàlana feno ho an'ny mpampiasa ho an'ny APEX VAC Pump, amin'ny antsipiriany momba ny fampandehanana, fiarovana, fepetra ara-teknika, ary famahana olana ho an'ny modely VacPro, VacMaxi, ary VacPlus avy amin'ny Apex Medical Corp.
Preview APEX XT Auto CPAP System Boky Torolàlana
Boky torolalana feno ho an'ny APEX XT Auto CPAP System (Model 9S-005720), amin'ny antsipiriany ny fametrahana, ny fampandehanana, ny fikojakojana, ny famahana olana, ary ny fepetra ara-teknika amin'ny fitsaboana Obstructive Sleep Apnea (OSA).
Preview APEX-V2 Massage Chair Instruction Manual - Setup, Operation, and Maintenance Guide
Comprehensive instruction manual for the APEX-V2 massage chair, covering safety precautions, installation, operation, settings, specifications, troubleshooting, cleaning, and maintenance. Learn how to set up and use your APEX-V2 massage chair effectively.
Preview DESTINY Security System Owner's Manual
Ity boky torolalana ho an'ny tompony ho an'ny Apex DESTINY Security System dia manome torolàlana feno momba ny fametrahana, ny fampandehanana, ny fomba fiadiana (AWAY, HOME, NIGHT), ny fidirana an-tanety, ny fiasan'ny passage, ary ny endri-javatra automatique an-trano.
Preview Apex AP-Pomp Massage Chair User Operation Manual
Comprehensive user operation manual for the Apex AP-Pomp massage chair, covering safety precautions, maintenance, installation, controller functions, troubleshooting, and specifications.