Extech PH90

Extech PH90 Waterproof pH Meter Instruction Manual

Modely: PH90

1. Vokatra mihoatraview

The Extech PH90 Waterproof pH Meter is a precision instrument designed for accurate pH measurement across various applications, including water quality testing, hydroponics, aquaculture, swimming pools, beverage production, and laboratory use. This device features a robust waterproof housing, a replaceable electrode, 3-point automatic calibration, and auto buffer recognition for ease of use and reliable results. Its compact design makes it suitable for both field and professional environments.

Extech PH90 Waterproof pH Meter

Figure 1: Extech PH90 Waterproof pH Meter. This image shows the compact, handheld pH meter with its digital display and control buttons, ready for use.

2. Endri-javatra fototra

3. Fametrahana sy Fiomanana

3.1. Fametrahana bateria

The PH90 meter requires two CR2032 Lithium Metal batteries (included). To install or replace batteries:

  1. Unscrew the battery compartment cap located at the top of the meter.
  2. Insert the two CR2032 batteries, ensuring correct polarity (+/-).
  3. Securely screw the battery compartment cap back into place.

3.2. Electrode Preparation

Before first use, and after storage, prepare the electrode:

  1. Remove the protective cap from the electrode tip. The cap contains a moist sponge to keep the electrode hydrated.
  2. Rinse the electrode tip thoroughly with distilled or deionized water. Do not wipe the electrode, as this can create static charges.
  3. Allow any excess water to drip off.

3.3. Calibration voalohany

For optimal accuracy, perform a 3-point calibration before initial use and periodically thereafter. Calibration buffers (pH 4, 7, and 10) are sold separately.

  1. Turn on the meter by pressing the ON/OFF bokotra.
  2. Immerse the electrode into a pH 7.00 buffer solution.
  3. Tsindrio ary hazòny ny kal button until "CAL" appears on the display. The meter will automatically recognize the buffer and calibrate.
  4. Sasao amin'ny rano voadio ny electrode.
  5. Repeat the process with pH 4.00 and pH 10.00 buffer solutions. The meter will automatically detect and calibrate to these points.
  6. Once calibration is complete, the meter will return to measurement mode.
Extech PH90 Meter Display and Buttons

Figure 2: Close-up of the Extech PH90 display and control buttons, including ON/OFF, MODE, and CAL buttons.

4. Torolàlana miasa

4.1. Maka refy

  1. Ensure the meter is calibrated.
  2. Rinse the electrode with distilled water and gently shake off excess.
  3. Immerse the electrode tip into the sample solution, ensuring the sensor is fully submerged.
  4. Gently stir the solution to ensure homogeneity and allow the reading to stabilize. This may take up to a minute, especially if the sample temperature differs significantly from the meter's previous environment.
  5. Read the pH value displayed on the LCD screen. The temperature will also be displayed.
  6. After measurement, rinse the electrode thoroughly with distilled water.

4.2. Asa fihazonana

To freeze the current reading on the display, press the MANOHANA button (if available, typically integrated with MODE or a separate button). Press it again to release the hold and return to live measurement.

4.3. Automatic Temperature Compensation (ATC)

The Extech PH90 features Automatic Temperature Compensation, which adjusts pH readings based on the sample's temperature, ensuring accuracy across varying temperatures. The temperature is displayed alongside the pH value.

5. Fikojakojana

5.1. Electrode Care and Storage

Proper electrode care is crucial for maintaining accuracy and extending the life of your meter:

Extech PH90 Electrode Tip

Figure 3: Close-up of the Extech PH90's replaceable electrode tip, which requires proper care and storage.

5.2. Fanoloana Electrode

The PH90 features a replaceable electrode. If the electrode performance degrades significantly despite proper cleaning and calibration, it may need replacement. Refer to the Extech website or contact customer support for information on purchasing replacement electrodes (Part PH95).

5.3. Fanoloana bateria

When the battery indicator appears on the display or the meter fails to power on, replace the two CR2032 Lithium Metal batteries as described in Section 3.1.

6. Famahana olana

OLANAAntony mety hitrangavahaolana
Famakiana tsy marina na tsy marin-toerana
  • Electrode not calibrated
  • Dirty or dry electrode
  • Temperature not equilibrated
  • Expired buffer solutions
  • Electrode simba
  • Perform a 3-point calibration.
  • Clean the electrode and ensure it is properly stored.
  • Allow sufficient time for the meter's temperature sensor to equilibrate with the sample.
  • Use fresh calibration buffer solutions.
  • Soloy ny electrode.
Tsy mandeha ny metatra
  • Batterie maty
  • Fametrahana bateria diso
  • Soloy amin'ny sela CR2032 vaovao ny bateria.
  • Jereo ny polarity bateria.
"CAL" error during calibration
  • Vahaolana buffer diso ampiasaina
  • Contaminated buffer solution
  • Electrode issue
  • Ensure correct pH buffer (4, 7, or 10) is used.
  • Use fresh, uncontaminated buffer solutions.
  • Clean or replace the electrode.

7. Famaritana

fikiranasarobidy
Laharana modelyPH90
pH Range0 hatramin'ny 14 pH
pH Accuracy± 0.01 pH
Temperature Range(Not explicitly stated, but integrated for ATC)
Loharanon-kery2 x CR2032 Lithium Metal batteries
Lanja entana2.88 ounces (81.6 grama eo ho eo)
Ny refin'ny vokatra6.7 x 1.4 x 1.4 santimetatra (tokony ho 17 x 3.6 x 3.6 cm)
Anisan'ny singaFlat surface pH electrode, cap, batteries (pH buffers sold separately)
mpanamboatraExtech

8. Warranty sy fanohanana

For warranty information, technical support, or to purchase replacement parts and accessories, please visit the official Extech website or contact Extech customer service. You can also find more information and products at the Fivarotana Extech ao amin'ny Amazon.

Documents mifandraika - PH90

Preview Extech ExStik™ Waterproof pH Meters PH100 & PH110 User Guide
Comprehensive user guide for the Extech ExStik™ Waterproof pH Meters, models PH100 and PH110. Learn about operation, calibration, maintenance, and specifications.
Preview Extech PH220 Waterproof Palm pH Meter User Manual
User manual for the Extech PH220 Waterproof Palm pH Meter. Covers introduction, meter description, operation, calibration, measurement, storage, battery replacement, maintenance, troubleshooting, and specifications.
Preview Extech MO257 Pinless Moisture Meter User Guide
User guide for the Extech MO257 Pinless Moisture Meter, detailing its features, operation, specifications, and maintenance for accurate non-invasive moisture measurements in building materials.
Preview Boky Torolàlana ho an'ny Mpampiasa ny Extech ET40 Heavy Duty Continuity Tester
Boky torolalana ho an'ny mpampiasa ny Extech ET40 Heavy Duty Continuity Tester. Mianara ny fomba fitsapana azo antoka sy mahomby ny fitohizan'ny singa tsy misy herinaratra, fuses, switches, relays, tariby ary circuit boards.
Preview Extech Electromagnetic Field Meter Model 480823 User Manual
User manual for the Extech Electromagnetic Field Meter, Model 480823. This document details the meter's specifications, operation, battery replacement, EMF exposure information, warranty, and calibration services. It measures EMF in Gauss and Tesla units with a frequency bandwidth of 30 to 300Hz.
Preview Extech ExStik Waterproof pH Meters PH100 & PH110 User Manual
Comprehensive user manual for the Extech ExStik Waterproof pH Meters, models PH100 and PH110. Provides detailed instructions on operation, calibration, electrode maintenance, battery replacement, safety precautions, specifications, and troubleshooting for these high-accuracy pH testing devices.